Every six months, Clean exterior surfaces, Calibration and system controls – Luminex 100 IS Version 2.2 User Manual

Page 181: Luminex 100 analyzer air intake filter

background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-069 Rev. A

7 - 5

Clean Exterior Surfaces

‹

To clean exterior surfaces, follow these steps:

1. Disconnect the system from ac power by turning off the power

switches and unplugging the Luminex 100 analyzer and the
Luminex XYP instrument.

2. Wipe all exterior surfaces with mild detergent, followed by a

10% bleach solution.

3. Open both doors of the system and clean all accessible surfaces

with detergent followed by a 10% bleach solution.

4. Dry the sheet metal surfaces to prevent corrosion.

5. Plug in and power on the Luminex 100 analyzer and the

Luminex XYP instrument.

6. Calibrate the system as outline in “Calibrate the System” on

page 5-2.

Calibration and System
Controls

Run calibration and system controls at least once a month during
routine use and:

following installation

if the system is powered off

if the system is moved

if a part is replaced

if the delta calibration temperature shown on the system monitor
(located on the Diagnostics tab) is more than ±3 degrees.

Each step usually takes less than one minute. You must run xMAP
controls after each calibration
. See “Run System xMAP Controls”
on page 5-4.

Every Six Months

Luminex 100 Analyzer Air
Intake Filter

‹

To replace the air intake filter:

1. Disconnect the Luminex 100 analyzer from ac power by turning

off the power switch on the rear of the analyzer, then unplugging
the power cord from the wall source.

2. On the back of the Luminex 100 analyzer, in the upper left

corner, remove the screw at the top of the panel and open the
panel door.

Note:

Hold on to the tubing! Do

not allow the tubing to fall inside t
the analyzer.