Assay lot management – Luminex 100 IS Version 2.2 User Manual

Page 129

background image

x

MAP Technology

Using Luminex 100 IS 2.2 Software

PN 89-00002-00-069 Rev. A

6 - 75

4. The kit manufacturer’s template appears on the selection list.

Click the name of the template.

5. Click Open. The template is now loaded and available for use.

Assay Lot Management

You can edit standard and control lot information. Once a lot is used,
changing or modifying it will prompt you for a new lot name. This
includes batches that have been setup, but not yet acquired. Lot
number information is not entered until it is used.

You can modify known lot concentration values. If you change the
known concentration for a used lot, the system prompts you to enter
a new lot name. If you change the lot concentration values for an
unused lot, the system updates the lot with the new lot concentration
values.

Once you import a template, lot information for the standards and
control reagents, specified in the template, must be entered. This lot
information is used for every batch setup using this template until
changed.

For assay reagents specified in templates the system allows you to
create new lots, edit lot information, select preexisting lots for reuse,
or import and export lots.

When editing lot numbers follow these lot handling rules:

1. If you have entered lot information for a template, but you have

not used the template to setup a batch, then you can edit the lot
value.

2. If you have entered lot information and you have used the

template to setup a batch (even if it has not yet been acquired),
then you can edit the lot values. However, you must rename the
lot.

Lot procedures include:

Create New Lots

Edit Lot Information on an Unused Template

Edit Lot Information on a Used Template

Import a Lot to Existing Template

Export Lot from Existing Template

Edit Lot Information