beautypg.com

Luminex 100 IS Version 2.2 User Manual

Page 48

background image

Luminex 100 IS User Manual Version 2.2

x

MAP Technology

5 - 4

PN 89-00002-00-069 Rev. A

15. Enter the expiration date.

16. Enter the values, listed in the Certificate of Quality that is

included with your calibrators, into the CAL2 entry fields.

17. Click OK.

18. Ensure that the analyzer is set to draw the CAL1 and CAL2

beads from the wells you loaded in step 2.

19. Click CAL1 from the Maintenance tab. A Confirmation Screen

dialog box appears.

20. Click OK and wait until CAL1 completes. The device status

section in the status bar changes from “Running” to “Standby”.

21. Click CAL2 from the Maintenance tab. A Confirmation Screen

dialog box appears.

22. Click OK and wait until CAL2 completes. The Device Status

section in the status bar changes from “Running” to “Standby”
and the Diagnostics tab turns red.

23. The System Monitor on the Diagnostics tab displays the date and

time in green if CAL1 and CAL2 are successful.

You must run system controls following calibration. Continue with
the following “Run System xMAP Controls” section.

Run System xMAP

Controls

To calibrate your system with controls:

1. Click New CON Targ. from the Maintenance tab to enter or

confirm system control lot numbers. An Update CON Target
Information dialog box appears. See Figure 10.

2. Enter the CON1 lot number.

3. Enter the expiration date. See step 11 under “Run System xMAP

Calibrators” on page 5-2 for detailed instructions

4. Enter the values listed from the Certificate of Quality included

with your system controls into the CON1 entry fields. You can
also find the target value information on the Luminex website at

http://luminexcorp.custhelp.com

.

5. Enter the CON2 lot number.

Note:

Ensure that you enter the

correct target values for each
parameter before clicking OK.
You can re-enter any incorrect
information.

Note:

The Diagnostics tab turns

red because all substances (CAL
or CON) have not been run and
thus occurs under the following
conditions:

• The first time the software

is opened

• The first time a system is

calibrated

• A new database is

installed

• An old database is

restored

• A CAL or CON fails.