Errors Tab

The Errors tab displays a list of errors that occurred during
acquisition. See Figure 90. It organizes the errors in two categories:

System and assay errors (top section of tab)

Sample errors (bottom section of tab)

Table 6 lists the errors in each category.

Table 6. Error Categories

System and Assay Errors

Sample Errors

Instrument Not Calibrated

Failed Verification (system
lists each failed control)

Failed Control (verifier
failed)

Temperature Divergence
from Calibration
Temperature

Failed Curve Fit

Unknown Diagnostic Error

APD Temp Range
Exceeded

XYP Temp Unstable

Low Voltage

High Sheath Pressure

Low Sheath Pressure

Command Timeout

Low Laser Power

Unknown Formula Failure.

Insufficient Bead Count

Sample High/Low

Diagnostic Error

High Sheath Pressure

Low Sheath Pressure

Sample Timeout

Sample Empty