Calibration and verification – Luminex 100 IS Version 2.2 User Manual
Page 87
x
MAP Technology
Using Luminex 100 IS 2.2 Software
PN 89-00002-00-069 Rev. A
6 - 33
•
Actuator test
•
Syringe pump test
•
Backflush valve test
•
Debubbler Valve test
To run self diagnostics:
1. Click Self-Diagnostics on the Maintenance tab. A confirmation
dialog box appears.
2. Click OK. The system begins processing the various self
diagnostic tests. As it completes the tests, the Status Bar changes
from a “Processing” command state to an “Idle” command state.
The self-diagnostic tests should take less than one minute to
complete.
3. The Command List on the Run Batch Tab indicates that Self-
Diagnostics is being performed. The Status Bar indicates that the
system is self diagnosing. Upon completion, the command list
indicates whether the command succeeded with green text or
command failed with red text.
If the self diagnosis fails you can obtain detailed information
regarding the results of the self-diagnostic test. See the following
“View Self-Diagnostic Details” section.
View Self-
Diagnostic Details
To view details of the self-diagnostics test that passed or failed:
1. Select the Diagnostics tab and view the Message Log. At least
one error from a failed self-diagnostics test appears with a
yellow background.
2. Double-click the yellow row to see a detailed description. An
Errors dialog box appears showing a list of passed and failed
self-diagnostic tests. Click OK to close this dialog box.
See the Troubleshooting chapter in this manual for additional
information about the self-diagnostic failure.
Calibration and
Verification
You can run calibration and verification commands from the
Maintenance tab.