Batches – Luminex 100 IS Version 2.2 User Manual
Page 108
Luminex 100 IS User Manual Version 2.2
x
MAP Technology
6 - 54
PN 89-00002-00-069 Rev. A
Analyze
Reprocessed Data
with Replay Batch
After running Replay Batch, you can analyze the data.
To analyze Replay Batch data:
1. Select Start Analysis on the toolbar.
2. At the Open Batch dialog box select the batch you want to
analyze and click Select. The most recent Replay Batch is the
last or has the highest ID number. See Figure 52.
Figure 52. Analyze Open Batch
3. The Analysis window appears presenting the Error tab,
Standards tab, and Sample tab data. See Figure 80 for an
example of the Standards tab. To close the Analysis window
click Close. See “Analyze Batches and Multi-Batches” on
page 6-86.
Batches
A batch consists of a group of samples processed under control of a
template. Batches are set-up using templates defined by assay kit
manufacturers.
Batches consist of templates and samples for acquisition, and can
span more than one plate. Templates contain predefined commands
that must be included in every batch acquisition. Commands may
consist of Acquire Standard, Acquire Control, Acquire Background,
Wash from Well, or Prime.
You can group batches together as a multi-batch. Multi-batches can
consist of any number of batches that have been setup from different
assay templates and are processed consecutively.
Note:
The Open Batch
dialog box does not list or
show the templates
associated to the batches.