Luminex 100 IS Version 2.2 User Manual

Page 47

background image

x

MAP

Technology

System Calibration

PN 89-00002-00-069 Rev. A

5 - 3

Click the Select Calendar button to display the calendar.

Use the arrows at the top left and right side of the calendar to
select past or future months. Alternately, select the month in
header. To select the year, click the displayed year in the header.
Use the up and down arrows that appear to select a past or future
year. Once you select the correct month and year, click the
desired expiration day on the calendar. The calendar closes and
the new month and year appear in the Expiration Date field.

Figure 9. Update Calibration Targets Dialog Box

13. Enter the values listed from the Certificates of Quality (COQ)

included with your calibrators into the CAL1 entry fields. You
can also find the target value information on the Luminex
website at

http://luminexcorp.custhelp.com

.

14. Enter the CAL2 lot number.