Luminex 100 IS Version 2.2 User Manual
Page 213
x
MAP Technology
Installation
PN 89-00002-00-069 Rev. A
A - 7
3. Remove the Cheminert fitting by turning the connector
counterclockwise. If the tubing is difficult to remove, push up
gently on the Sample Probe. See Figure 137.
Figure 137. Cheminert Fitting
4. Remove the Sample Probe by grasping the probe and gently
pushing up. Read the Caution to the left.
5. Insert the Luminex XYP instrument Sample Probe and replace
the tubing connector. Tighten the connector by turning the
connector clockwise, being careful that the threads are correctly
aligned. Hand tighten only See Figure 138.
1. Cheminert Fitting
2. Sample Probe
1
2
Caution:
The Luminex XYP instrument Sample Probe should
slide up easily while removing it from the Sample Arm. If you feel
resistance, do not force the probe up. Contact Luminex
Customer Support for assistance.