Luminex 100 IS User Manual Version 2.2

x

MAP Technology

6 - 24

PN 89-00002-00-069 Rev. A

Maintenance Tab

At the Maintenance tab you perform maintenance commands, XYP
commands, and calibration and verification commands during system
use. See Figure 35.

Figure 35. Maintenance Tab

Maintenance Commands

Perform these commands to maintain your system and to keep it in
optimal working condition. Follow the step-by-step instructions for
each command (found after this section) when using as-needed
maintenance operations. Some of the commands are also available on
the Home tab. The available commands are:

Warmup

Prime

Backflush

Alcohol Flush - must use the Luminex XYP reservoir due to
volume requirement

Sanitize - must use the Luminex XYP reservoir due to volume
requirement

Wash

Drain

Soak

Self Diagnostics

When the system processes a command that you select from the
Maintenance tab, an animated gear graphic appears for a short time
in the upper-right corner of the Maintenance tab. The system labels

Note:

The Sanitize command

performs a similar function as
the alcohol flush command.
However, the sanitize
command uses 10% to 20%
household bleach to
decontaminate sample lines
and the cuvette after
biohazard contact.