Luminex 100 IS User Manual Version 2.2

x

MAP Technology

6 - 6

PN 89-00002-00-069 Rev. A

batches. The automatic analysis feature takes place after the system
finishes acquiring data while processing batches.

‹

To configure General tab information:

1. Click Tools from the menu bar and select Options. The Options

dialog box appears. See Figure 20.

2. Click the General tab.

3. Select the browse button

and navigate to a different

directory (folder) if you want the batch files written to a
directory other than

C:\My Batches

(the default location).

4. If desired, enter the name of the current user or operator in the

Current User field.

5. Enter the number of digits that you want to appear in your

sample data in the Analysis Display Digits field.

Figure 20. Options Dialog Box—General Tab

6. Select the desired checkboxes to enable:

• Display Confirmation Screens
• Enable Raw Data Storage
• Report Raw Flourescence
• Auto-start Analysis

7. When satisfied with your selections click OK. The system saves

the General tab information to the database.

Note:

You cannot select both

Auto-start Analysis and Auto
Export Batches (Data Export
tab).