Luminex 100 IS Version 2.2 User Manual
Page 60

Luminex 100 IS User Manual Version 2.2
x
MAP Technology
6 - 6
PN 89-00002-00-069 Rev. A
batches. The automatic analysis feature takes place after the system
finishes acquiring data while processing batches.
To configure General tab information:
1. Click Tools from the menu bar and select Options. The Options
dialog box appears. See Figure 20.
2. Click the General tab.
3. Select the browse button
and navigate to a different
directory (folder) if you want the batch files written to a
directory other than
C:\My Batches
(the default location).
4. If desired, enter the name of the current user or operator in the
Current User field.
5. Enter the number of digits that you want to appear in your
sample data in the Analysis Display Digits field.
Figure 20. Options Dialog Box—General Tab
6. Select the desired checkboxes to enable:
• Display Confirmation Screens
• Enable Raw Data Storage
• Report Raw Flourescence
• Auto-start Analysis
7. When satisfied with your selections click OK. The system saves
the General tab information to the database.
Note:
You cannot select both
Auto-start Analysis and Auto
Export Batches (Data Export
tab).