Luminex 100 IS User Manual Version 2.2

x

MAP Technology

6 - 18

PN 89-00002-00-069 Rev. A

Data Acquisition
Categories

The Home tab contains categories representing different phases of
data acquisition. Each category consists of a group of command
buttons associated with typical phases of data acquisition. At the
click of a button, the system performs or initiates the process to
perform the function for the button you click. See Figure 26.

At the Daily Startup section you perform the recommended daily
startup procedure by selecting the buttons in the order they appear.
These buttons include warm up, prime, alcohol flush, and wash the
system (sheath fluid). See “Maintenance Commands” on page 6-24
for more detailed information regarding startup functions.

At the Instrument Calibration section you perform system
calibration and control functions. This section contains CAL1,
CAL2, CON1, and CON2 buttons.

At the Batch Setup section you create a new batch, create a new
multi-batch, and create new lots for an assay’s standards and
controls.

At the Reports and Analysis section you analyze the data acquired
from your batches, compile the data into reports, and print reports
from this analysis.

The Daily Shutdown section consists of two operations typically
recommended during daily shutdown procedures. This section lets
you sanitize (bleach) and soak (distilled water) the system.