beautypg.com

Luminex 100 IS Version 2.2 User Manual

Page 149

background image

x

MAP Technology

Using Luminex 100 IS 2.2 Software

PN 89-00002-00-069 Rev. A

6 - 95

Figure 85. Invalidate Data Analysis Example

Expected

Concentrations

The Standards tab displays an Expected Concentration column for
standard and control samples. The Standards Expected Concentration
column allows users to edit the standard concentrations in that field.

You can edit controls, however, you must select the control name or
Expected Concentration field for that control, then select the Change
Lot (Alt+F8)
button.

When editing standard and control lot values the system may prompt
you for a new lot number. If this is the first batch you analyze using
this lot number, the system allows editing without requiring a new lot
name. However, if you have analyzed a previous batch using this lot,
the system will require that you rename the lot if edited or modified.

Change Lots

Use this feature to edit the lot that is applied to the batch currently
opened in data analysis.

‹

To change the lot that is applied to a batch:

1. Click Change Lot (Alt + F8), located at the bottom of Analysis

window, to display the Choose Lot dialog box. See Figure 86.

2. The dialog box displays a list of available standard and control

lots that you can apply to a batch. Highlight the desired lot and
click OK to apply the selected lot to the batch opened in data
analysis. Figure 86 shows standard and control. Figure 87 shows
standard only.

Note:

Restrict editing lot values

for use only in correcting user
entry errors for standards and
control lot values.