Luminex 100 analyzer general, Optics, Fluidics – Luminex 100 IS Version 2.2 User Manual
Page 31: Electronics

x
MAP
Technology
The System
PN 89-00002-00-069 Rev. A
3 - 5
Luminex 100
Analyzer General
•
Operating temperature: 15°C to 30°C (59°F to 86°F)
•
Storage temperature: 0°C to 50°C (32°F to 122°F)
•
Humidity: 20% to 80%, noncondensing
•
Altitude: Operation up to 2400 m (7874 ft.) above mean sea level
•
Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H
•
Weight: maximum of 25 kg (60 lbs.)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
•
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
•
Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to + 50°C and 20-80%
noncondensing, respectively
•
Input voltage range: 100 - 240 V~, 1.5 Amp, 47-63 Hz or 100 -
120 V~, 1.4 Amp, 47-63 Hz. (refer to the serial number label on
the rear panel of the instrument)
•
AC inlet fuse: 3 Amp, 250 V~, fast acting or 2 Amp, 250 V~,
fast acting (refer to the fuse label on the rear panel of the
instrument)
Optics
•
Reporter laser: 532 nm, nominal output 10-16.5 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)
•
Classification laser: 635 nm, 9.1 mW ± 6%, maximum output 25
mW, diode; mode of operation, continuous wave (CW)
•
Reporter detector: Photomultiplier tube, detection bandwidth of
565-585 nm
•
Classification detector: Avalanche photo diodes with temperature
compensation
•
Doublet discrimination detector: Avalanche photo diodes with
temperature compensation
Fluidics
•
Sheath flow rate: 90 µL ± 5 µL/second
•
Cuvette: 200 micron square flow channel
•
Sample injection rate: 1 µL/second
•
Sample uptake volume: 20-200 µL
Electronics
•
Reporter channel detection: A/D resolution 14 bits
•
Communications interface: USB
•
Luminex XYP instrument, communications interface: RS 232