System error states – Luminex 100 IS Version 2.2 User Manual

Page 196

background image

Luminex 100 IS User Manual Version 2.2

x

MAP Technology

8 - 6

PN 89-00002-00-069 Rev. A

Low Voltage

Detected

Power off the Luminex 100 analyzer, Luminex XYP instrument,
and PC. Power on the system.

Calibrate and verify the system.

If the error message occurs a second time, contact Luminex
Customer Support.

Runtime Sheath

Pressure Out of

Limits (Too High)

Verify that the sheath fluid container is at the same level as the
analyzer.

On systems with a Luminex Sheath Delivery system, ensure that
the sheath fluid pressure reading is equal to the original reading
by adjusting the regulator as outlined in the Luminex Sheath
Delivery System
appendix.

Calibrate and verify the system.

Runtime Sheath

Pressure Out of

Limits (Too Low)

Verify that the sheath fluid container lines are attached to the
appropriate connectors on the Luminex 100 analyzer.

Verify that there is no sheath fluid or air leaking from the sheath
fluid container connections.

Check for sheath fluid leaks at all connections.

Calibrate and verify the system.

Low Laser Power

Detected

Warm up the system. This takes approximately 30 minutes.

Calibrate and verify the system.

Open the calibration trend report for L100-CAL2 and record the
last three reporter voltages. Contact Luminex Customer Support
with this information.

Calibrate and verify the system daily before use.

Command

Encountered Time

Out Error

This error message occurs when the current command fails. For
example, if the system suddenly lost power or if the system is
unable to pressurize in the set time. Repeat the command after
checking the system.

System Error States

The following states are displayed on the Status Bar or on the Run
Batch tab when the system is in operation.

Disconnected

Click the Connect button on the toolbar and wait for status to
change.

Remove and reconnect the cable connections.

Close the software application and reopen it.

Close the software application and power off the Luminex 100
IS 2.2 system (Luminex 100 analyzer, Luminex XYP instrument,
and PC). Power on the system and check status.