Luminex 100 IS User Manual Version 2.2

x

MAP Technology

6 - 64

PN 89-00002-00-069 Rev. A

Open a Multi-

Batch

Use this procedure to open a multi-batch that you created earlier for
acquisition.

‹

To load a multi-batch for acquisition:

1. Click Open Multi-Batch from the toolbar. An Open Multi-

Batch dialog box appears listing the available multi-batches for
selection.

2. Select any one of the batches having the desired multi-batch

name.

3. Click Select. The system loads the multi-batch to the Run Batch

tab.

4. Click Eject/Retract to eject the plate holder.

5. Load the first microtiter plate onto the plate holder.

6. Once you load the plate to the system, click Start Plate to retract

the plate holder and begin acquiring the multi-batch data.

Process Multiple

Plates

You can process multiple plates per batch or multi-batch as required.
After loading a batch or multi-batch that spans more than one plate, a
new plate appears to the immediate right of the existing plate image
on the screen. A dark blue line separates the adjacent columns of the
two plates. See Figure 60.

Figure 60. Processing Multiple Plates

Note:

Each batch within a

multi-batch appears on the list
of available multi-batches. The
software differentiates them
according to the batch ID,
name, and description.
However, the software gives all
batches comprising a multi-
batch the same multi-batch ID
number and name.