Luminex 100 IS User Manual Version 2.2
x
MAP Technology
6 - 92
PN 89-00002-00-069 Rev. A
Figure 82. Analysis Window—Standards Tab
Function Keys
On the lower portion of the Analysis window the system lists nine
function keys. These keys perform tasks or functions that are
relevant to the Standards tab although they appear on all three tabs of
the Analysis window. The nine function keys are:
•
Next Test (F2)
•
Previous Test (F3)
•
Invalidate Standard (F4)—see “Invalidate or Validate Standards
and Controls” on page 6-93
•
Validate Standard (F5)—see “Invalidate or Validate Standards
and Controls” on page 6-93
•
Invalidate Control (F6)— see “Invalidate or Validate Standards
and Controls” on page 6-93
•
Validate Control (F7)—see “Invalidate or Validate Standards
and Controls” on page 6-93
•
Change Lot (Alt + F8)—see “Change Lots” on page 6-95
•
Recalc (Standards tab only)—see “Recalculation” on page 6-93
Edit lots—Use the Change Lot (Alt+F8) button to edit lots. Restrict
the use of this button to correct only entry errors for standard and
control lot values. To create new lots refer to “Create a New Lot” on
page 6-76.