Luminex 100 IS Version 2.2 User Manual

Page 86

background image

Luminex 100 IS User Manual Version 2.2

x

MAP Technology

6 - 32

PN 89-00002-00-069 Rev. A

2. Click the drop-down arrow to the right of the Eject/Retract

button. An image of the microtiter plate and Luminex XYP
reservoir appear. See Figure 39.

Figure 39. Microtiter Plate and XYP Reservoir Image

3. Select a location on the image of the microtiter plate and

reservoir containing the draw fluid and then click Eject/Retract.
The plate holder ejects.

4. Verify that the location you indicated is empty, place the distilled

water in the location, and click Eject/Retract to retract the plate
holder.

5. Click OK to continue. The command list on the Run Batch tab

indicates that it is running. The Device Activity field on the
Status Bar indicates that the system is soaking. Wait until the
soak command completes before initiating another command.
Upon completion, the command list indicates whether the
command succeeded with green text or the command failed with
red text.

Self Diagnostics

Command

Run self diagnostics to see if the Luminex 100 analyzer and all
system operations are running correctly.

The system tests these functions when running self diagnostics:

Flash memory test

Microcontroller RAM test

Nonvolatile memory test

DSP program CRC test

DSP capture test

High voltage module

Channel backgrounds test

Pressurization test

Note:

The system may show

a default location. In this case,
the microtiter plate shows a
well with a yellow line inside
the well.

Note:

Running the self

diagnostics test resets the
laser and begins a 30-minute
warmup process.