beautypg.com

Luminex 100 IS Version 2.2 User Manual

Page 139

background image

x

MAP Technology

Using Luminex 100 IS 2.2 Software

PN 89-00002-00-069 Rev. A

6 - 85

Once you start processing a batch, you cannot reacquire data. If you
have used a batch for acquisition before, the system will not display
it in the open batch list.

Pause

The Pause command temporarily interrupts the acquisition process.
Some examples to pause the system are:

The temperature goes out of the designated range during
acquisition

To add sheath fluid

To add a new reagent to your sample group

‹

To pause a command:

Click Pause on the Run Batch tab or the Acquisition Detail tab.
The system completes the current command in process and then
pauses the system.

Resume

The Resume command continues an interrupted acquisition process.

‹

To resume a batch:

Click Resume on the Run Batch tab or the Acquisition Detail
tab. The system resumes processing the command list from
where it was paused.

Cancel Command

The Cancel Command button cancels the current command in
progress.

You cannot cancel the Warmup command.

‹

To cancel a command:

Click Cancel Cmd on the Run Batch tab or the Acquisition
Detail tab. The command cancels.

Cancel All

The Cancel All command cancels the entire batch.

‹

To cancel an acquisition once it has begun:

Click Cancel All on the Run Batch tab or the Acquisition Detail
tab. The current series of commands or batch acquisition is
cancelled.

Upon cancelling, the system stops its activity regardless of its status.
The system highlights the well in yellow indicating the ending
acquisition process.

Note:

Selecting Cancel All

stops batch acquisition and you
will not be able to process it
again. Therefore, ensure that you
want to cancel the current batch,
multi-batch, command series, or
operation.