Luminex 100 IS Version 2.2 User Manual
Page 198

Luminex 100 IS User Manual Version 2.2
x
MAP Technology
8 - 8
PN 89-00002-00-069 Rev. A
•
Verify that the delta cal temperature is within
±
3.0 degrees.
Recalibrate if outside the limits and retry.
•
If the system controls pass, contact the kit manufacturer.
Temperature
Divergence from
Calibration
Temperature
•
Calibrate and verify the system.
•
If this message still appears following calibration, contact
Luminex Customer Support.
Failed Control in
Batch
This message occurs when one or more of the assay controls are not
within the expected range.
•
Verify that you selected the correct template.
•
Verify that the correct assay control lot number and
concentration values were selected and entered correctly.
•
Verify that the system is functioning properly by running the
system controls.
•
If the system controls pass, contact the kit manufacturer.
Failed Curve Fit
This message occurs when the calculation could not occur for the
assay standards for data interpretation.
•
Verify that you selected the correct template.
•
Verify that the correct control lot number and target values were
selected and entered correctly.
•
Verify that the system is functioning properly by running the
system controls.
•
If the system controls pass, contact the kit manufacturer.
Sample High/Low
This message occurs if the sample results do not fall within the
highest and lowest standards.
•
Verify that you selected the correct template.
•
Verify that the correct standard lot number and target values
were selected and entered correctly.
•
Verify that the system is functioning properly by running the
system controls.
•
If the sample is noted as High, dilute the sample following the
kit manufacturer’s suggestions.
•
If the system controls pass, contact the kit manufacturer.
Note:
We recommend that
you install the system in a
temperature-controlled
environment.