Status bar – Luminex 100 IS Version 2.2 User Manual
Page 65
x
MAP Technology
Using Luminex 100 IS 2.2 Software
PN 89-00002-00-069 Rev. A
6 - 11
between every command or acquisition until you disable (or
deselect) the single step option.
Status Bar
The status bar appears along the bottom of the 100 IS Main window
and displays the system status. Check the status bar periodically
while running a batch to ensure that the system is performing
accurately. See Figure 25.
Figure 25. System Status Bar
The system displays information about the Command State, Device
Status, Device Activity, Laser Status, Total Events per Second, and
Region Events per Second. Color coding indicates the urgency of
each item’s status. Device Activity uses no color coding.
Table 1 describes the types of status bar message in relation to the
message color coding.
Table 1. Status Bar Color Coding
Category
Color
Indicates
Command
State
Indicates communication with the Luminex 100 analyzer or
operations being processed
Green
Idle or processing
Yellow
Connecting, pausing, or paused
Red
Disconnected or locked out
Device Status
Indicates the current process of or warning about the
Luminex 100 analyzer
Green
Running or standby
Yellow
Busy, pressurizing, sheath is empty, or warming
up
Red
Not ready or disconnected
Device
Activity
Indicates the activity that the Luminex 100 analyzer is
performing
Note:
The device activity has no color code.