Mainframe – AMETEK ReFlex User Manual
Page 47
Mainframe
ReFlex Power™
2-12
M380056-01 Rev J
parallel. The load that would be applied to a particular AC/DC input line is
dependent on the location of the modules in the Mainframe slots.
The slots are grouped in sets of four: Slot 1-4, Slot 5-8, and Slot 9-12. Slot 1 is
the left-most slot, as viewed from the front (refer to Figure 2-5. Each of the
backplane connectors of a set of four slots is connected in parallel and is
supplied by one pair of the AC/DC input lines. In a double-width and triple-width
module, the module connector is located so that it mates to a backplane
connector in the right-most slot, as viewed from the front of the Mainframe. Refer
to Figure 2-4 for a diagram showing the power distribution from the input
connector to the twelve Mainframe slots.
AC/DC
Input
L1
L1-RTN
L2
L2-RTN
L3
L3-RTN
GND
Slot 1 Power
Slot 2 Power
Slot 3 Power
Slot 4 Power
Slot 5 Power
Slot 6 Power
Slot 7 Power
Slot 8 Power
Slot 9 Power
Slot 10 Power
Slot 11 Power
Slot 12 Power
L1 Power to Slots 1-4
L2 Power to Slots 5-8
L3 Power to Slots 9-12
Safety
GND
Mainframe
Figure 2-4. Input Power Distribution (Slot 1 at top is left-most slot of Mainframe
front view, as in Figure 2-5).