2 recommended calibration equipment, 3 calibration setup, Power supply module – AMETEK ReFlex User Manual
Page 269: Recommended, Calibration, Equipment, Setup
DC Power Supplies
ReFlex Power™
6-66
M380056-01 Rev K
6.6.2
RECOMMENDED
CALIBRATION
EQUIPMENT
The recommended calibration equipment is listed in Table 5-14. Equivalent
substitutes are acceptable provided that their accuracy is at least four times
better than the accuracy of the parameter specification that is being calibrated. If
less accurate equipment is utilized, measurement uncertainty may be introduced
that would compromise the validity of the calibration.
Table 6-15. Recommended Calibration Equipment
INSTRUMENT
REQUIREMENTS
RECOMMENDED
MODEL
DC Voltmeter
Resolution: 6 1/2 digits
Agilent 34401A
Current Shunt
RFP-D1016-021-XXXX, RFP-D2033-030-XXXX, RFP-
D2050-020-XXXX, and RFP-D2050-025-XXXX:
10mΩ, ±0.04%, (100A)
RFP-D1065-5A1-XXXX, RFP-D2120-8A3-XXXX, and
RFP-D2450-2A3-XXXX:
50mΩ, ±0.04%, (15A)
Guildline 9230-100
Guildline 9230-15
6.6.3
CALIBRATION
SETUP
Calibration must be performed under controlled environmental conditions. The
ambient temperature must be between 20 to 30 °C. Also, the humidity must be
less than 80%. Allow the unit to warm up for 30 minutes at no load prior to
performing the calibration. Refer to Figure 6-23 for a diagram of the test
instrumentation setup.
(-)
(-)
(+)
(+)
Output
Remote Sense
Power
Supply
Module
DVM
Current
Shunt
Closed for
current
calibration