2 recommended calibration equipment, 3 calibration setup, Power supply module – AMETEK ReFlex User Manual

Page 269: Recommended, Calibration, Equipment, Setup

background image

DC Power Supplies

ReFlex Power™

6-66

M380056-01 Rev K

6.6.2

RECOMMENDED

CALIBRATION

EQUIPMENT

The recommended calibration equipment is listed in Table 5-14. Equivalent
substitutes are acceptable provided that their accuracy is at least four times
better than the accuracy of the parameter specification that is being calibrated. If
less accurate equipment is utilized, measurement uncertainty may be introduced
that would compromise the validity of the calibration.

Table 6-15. Recommended Calibration Equipment

INSTRUMENT

REQUIREMENTS

RECOMMENDED

MODEL

DC Voltmeter

Resolution: 6 1/2 digits

Agilent 34401A

Current Shunt

RFP-D1016-021-XXXX, RFP-D2033-030-XXXX, RFP-
D2050-020-XXXX, and RFP-D2050-025-XXXX:
10mΩ, ±0.04%, (100A)
RFP-D1065-5A1-XXXX, RFP-D2120-8A3-XXXX, and
RFP-D2450-2A3-XXXX:
50mΩ, ±0.04%, (15A)

Guildline 9230-100

Guildline 9230-15

6.6.3

CALIBRATION

SETUP

Calibration must be performed under controlled environmental conditions. The
ambient temperature must be between 20 to 30 °C. Also, the humidity must be
less than 80%. Allow the unit to warm up for 30 minutes at no load prior to
performing the calibration. Refer to Figure 6-23 for a diagram of the test
instrumentation setup.

(-)

(-)

(+)

(+)

Output

Remote Sense

Power
Supply
Module

DVM

Current
Shunt

Closed for
current
calibration