14 hpal resistance calibration – part ii of v, Hpal – AMETEK ReFlex User Manual
Page 320
ReFlex Power™
Active Loads
M380056-01 Rev J
7-45
7.6.14
HPAL
R
ESISTANCE
C
ALIBRATION
–
P
ART
II
OF
V
This procedure calibrates the resistance for voltage gain 4 and current
range high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Connect the
sense positive-to-positive lead and sense negative-to-negative lead. Also
connect the shunt and the DC current meter between the active load and
DC power supply
This procedure requires maximum voltage 20.0 V and maximum current
3.0Amps.
1. Set the DC input voltage to 20 Volts
2. Turn the active load on:
INP
3. Set the conductance count to 0x1FFFF.
CAL
4. After DC voltage stabilizes. Enter the voltage using SCPI
commands
CAL
5. After DC current stabilizes. Enter the current using SCPI commands
CAL
6. Set the conductance count to 0xF0F0.
CAL