beautypg.com

4 output voltage parameter calibration, Output, Voltage – AMETEK ReFlex User Manual

Page 270: Parameter, Calibration

background image

ReFlex Power™

DC Power Supplies

M380056-01 Rev K

6-67

Figure 6-23. Calibration Instrumentation Setup

6.6.4

OUTPUT

VOLTAGE

PARAMETER

CALIBRATION

The following procedure calibrates parameters associated with the output voltage:
programming voltage (remote sense), readback voltage, overvoltage protection (OVP),
undervoltage protection (UVP), and readback of local voltage (internally sensed).
Voltage calibration is made with remote sensing, and with external sense leads
connected to the front panel output terminals. Voltage measurements with the external
voltmeter are made at the front panel output terminals, at the same points where the
sense leads are connected.

1. Connect a voltmeter across the output terminals. Ensure that no load is

connected to the output.

2. Issue clear and reset commands:


*CLS[n]

*RST[n]

3. Set the OVP supervisory to non-tracking and at maximum range:


CAL[n]:OUTP:VOLT:PROT:COUNTS 4095

4. Turn on the output:

OUTP[n]:ISOL on

OUTP[n]:SENS on

OUTP[n]:STAT on

5. Set the output current to full-scale:


CAL[n]:OUTP:CURR:COUNTS 32768

6. Program the output voltage to the first value near 5% of full-scale,

(count = 48852*0.05+768 = 3210):

CAL[n]:OUTP:VOLT:COUNTS 3210

7. Measure the output voltage with the external voltmeter. Enter this as value-1 with

the following command:

CAL[n]:OUTP:VOLT:FIVEPOINT 1

8. Program the output voltage to the second value near 27.5% of full-scale,

(count = 48852*0.275+768 = 14202):

This manual is related to the following products: