AMETEK ReFlex User Manual
Page 153
AC Power Supplies
ReFlex Power™
4-68
M380056-01 Rev K
14. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
15. Program the output current to 50% of the full scale current:
CAL
16. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
17. Program the output current to 72.5% of the full scale current:
CAL
18. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
19. Program the output current to 95% of the full scale current:
CAL
20. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
21. Turn the unit off:
OUTP
22. Perform the current calibration:
CAL
23. Send the system error query command to make sure calibration was successful:
SYST:ERR?
24. If there is no error returned at the system error query command then save the
calibration results:
CAL
CAL
CAL
25. Optional Commands: