beautypg.com

AMETEK ReFlex User Manual

Page 158

background image

ReFlex Power™

AC Power Supplies

M380056-01 Rev K

4-73

4.6.20

H

ALL

C

ALIBRATION FOR

V

OLTAGE

R

ANGE

L

OW

(5

KH

Z

U

NIT

)

The following procedure calibrates the hall gain sensor of the power supply for
the 5KHz AC unit only. This is seven-point calibration procedure.

1. Close the switch between the power supply and the resistive load.

2. Connect the meter between the output and adjustable resistive load.

3. Set the voltmeter to read the AC current.

4. Ensure that a resistive load is connected to the output.

5. Issue clear and reset commands:

*CLS
*RST

6. Set the voltage range of the power supply to low:

SOUR:VOLT:RANGE 0

7. Turn the unit on:

OUTP:STATE 1

8. Set the hall calibration data to the default value:

CAL:OUTP:FREQ:CURR:DEFAULT

9. Set the output voltage to 120 Volts:

SOUR:VOLT 120

10. Set the output current to 6.3 Amps:

SOUR:CURR 6.3

11. Turn on the resistive load and adjust the load until the output voltage is 110 Volts,

this will ensure that the AC supply is in Constant Current Mode.

12. Set the output frequency to 60Hz

SOUR:FREQ 60

13. Wait for AC current to stabilize. Measure the output current using the AC meter.

Enter the measured current:

CAL:OUTP:FREQ:CURR1

14. Set the output frequency to 500Hz

SOUR:FREQ 500