7.6.9

HPAL

V

OLTAGE

C

ALIBRATION

P

ART

II

OF

V

This procedure calibrates the voltage for voltage gain 4 and current range high.
Make sure that the shunt and the DC voltmeter are connected between the
positive and negative lead of the active load. Also connect the sense positive-to-
positive lead and sense negative-to-negative lead.

This procedure requires maximum voltage 109V and maximum current 6.0
Amps.

1. Turn the active load on:

INP:STATE 1

2. Set the conductance count to 0xCCCC:

CAL:INP:COND:COUNTS 0xCCCC

3. Set the input DC voltage to 27 Volts. After the input DC voltage stabilizes

enter the first exact voltage:

CAL:INP:VOLT 1

4. Set the input DC voltage to 109 Volts. After the input DC voltage

stabilizes enter the second exact voltage:

CAL:INP:VOLT 2

5. Send the calibration command to calculate the gain and offset:

CAL:MEAS:OFFSET:CALC

6. Send the system error query command to make sure calibration was

successful:

SYST:ERR?

7. If there is no error returned at the system error query command then save

the calibration result as follows:

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

8. Reset the unit

*RST