5 calpot calibration voltage range high – AMETEK ReFlex User Manual

Page 133

background image

AC Power Supplies

ReFlex Power™

4-48

M380056-01 Rev K

4.6.5

C

AL

P

OT

C

ALIBRATION

V

OLTAGE

R

ANGE

H

IGH

This procedure calibrates cal-pot for the voltage range high. This calibration is
done with the AC current meter connected to the output lead of the AC power
supply.

1. Connect the AC current meter to the output lead of the power supply. Ensure that

no load is connected to the output.

2. Issue clear and reset commands:

*CLS
*RST

3. Set the voltage range of the power supply to high

SOUR:VOLT:RANGE 1

4. Start the internal calibration of the offset pot

CAL:CALPOTS:OFFSET:CALC

5. Start the calibration of the gain pot

CAL:CALPOTS:GAIN:START

6. Set the load box to approximately 3.25 Amps. Wait for the current to stabilize

then make the AC current measurement using the meter connected to the power
supplies output lead.

CAL:CALPOTS:GAIN:CURR

7. Finish the gain pot calibration

CAL:CALPOTS:GAIN:CALC

8. Send the system error query command to make sure calibration was successful.

SYST:ERR?

9. If there is no error returned at the system error query command then save the

calibration result with the following commands

CAL:UNLOCK “6867”
CAL:STORE
CAL:LOCK

10. Optional Commands:

To query the offset pot set value: CAL:CALPOTS:OFFSET?