AMETEK ReFlex User Manual
Page 141

AC Power Supplies
ReFlex Power™
4-56
M380056-01 Rev K
4.6.10
C
URRENT
ADC
O
FFSET
C
ALIBRATION FOR
V
OLTAGE
R
ANGE
L
OW
The following procedure calibrates offset of the current ADC for the voltage range
low.
1. Ensure that no load is connected to the output.
2. Issue clear and reset commands:
*CLS
*RST
3. Set the voltage range of the power supply to low
SOUR
4. Perform the voltage ADC offset calibration
CAL
5. Send the system error query command to make sure calibration was successful.
SYST:ERR?
6. If there is no error returned at the system error query command then save the
calibration result with the following commands
CAL
CAL
CAL
7. Optional Commands:
To query the current ADC offset:
CAL
8. Reset the unit
*RST