4 default operational conditions, Power-on conditions, Reset conditions – AMETEK ReFlex User Manual
Page 302
ReFlex Power™
Active Loads
M380056-01 Rev J
7-27
Figure 7-9. Standalone Input Configuration
7.5.4
D
EFAULT
O
PERATIONAL
C
ONDITIONS
P
OWER
-O
N
C
ONDITIONS
When an Active Load module is first powered on, the following parameters
are set to factory default values:
•
Input configuration: standalone
•
Input current: 0A
•
Overvoltage protection (OVP): 110% of full-scale voltage; programmable mode
•
over power protection (OPP): 105% of the Load/s power rating
•
Overcurrent protection (OCP): 120% of full-scale current; programmable mode
•
Input state: off
•
Input isolation and remote sense relays: open
R
ESET
C
ONDITIONS
If a module is reset with the SCPI command, *RST[n], (where n is the module
address), the following default conditions are set:
•
Input configuration: unchanged
•
Input current: 0A
•
Overvoltage protection (OVP): 110% of full-scale voltage; programmable mode
•
Undervoltage protection (UVP): 0V
•
Overcurrent protection (OCP): 120% of full-scale current; programmable mode
•
Input state: off
•
Input isolation and remote sense relays: open (if present)
•
All status reporting data structures are cleared. Enable-masks are not cleared.