4 default operational conditions, Power-on conditions, Reset conditions – AMETEK ReFlex User Manual

Page 302

background image

ReFlex Power™

Active Loads

M380056-01 Rev J

7-27

Figure 7-9. Standalone Input Configuration

7.5.4

D

EFAULT

O

PERATIONAL

C

ONDITIONS

P

OWER

-O

N

C

ONDITIONS

When an Active Load module is first powered on, the following parameters
are set to factory default values:

Input configuration: standalone

Input current: 0A

Overvoltage protection (OVP): 110% of full-scale voltage; programmable mode

over power protection (OPP): 105% of the Load/s power rating

Overcurrent protection (OCP): 120% of full-scale current; programmable mode

Input state: off

Input isolation and remote sense relays: open

R

ESET

C

ONDITIONS

If a module is reset with the SCPI command, *RST[n], (where n is the module
address), the following default conditions are set:

Input configuration: unchanged

Input current: 0A

Overvoltage protection (OVP): 110% of full-scale voltage; programmable mode

Undervoltage protection (UVP): 0V

Overcurrent protection (OCP): 120% of full-scale current; programmable mode

Input state: off

Input isolation and remote sense relays: open (if present)

All status reporting data structures are cleared. Enable-masks are not cleared.