3 installation – AMETEK ReFlex User Manual

Page 102

background image

ReFlex Power™

AC Power Supplies

M380056-01 Rev K

4-17

4.4.3

I

NSTALLATION

CAUTION

The module contains ESD sensitive circuitry. Follow appropriate handling and grounding
procedures to prevent damage to the module. The module should be handled by the
enclosure, with care to protect the connectors from ESD discharges.


CAUTION

Input power must be disconnected to the Mainframe before a module is
inserted or extracted. Inserting or extracting a module with power installed (hot-
swapping) could result in damage to the module or Mainframe. Prior to turning
power on, ensure that front panel captive fasteners are tightened.

Modules are installed into the Mainframe from the front of the chassis. Refer to
Figure 4-4 for the orientation of a module and Mainframe during installation.

1. Ensure that the AC/DC input power is disconnected before

installing or removing any module(s).

2. Insert the module into the mainframe, rear panel first, aligning the

module guide rails the top and bottom Mainframe guides.

3. Gently slide the module completely into the mainframe until the

guide pins of the module rear connector engage the mating
backplane connector.

4. Fully seat the module until the front panel inner surface is snug

against the mainframe chassis.

5. Secure the module(s) to the Mainframe at the top and bottom of

each front panel by turning the captive thumbscrews clockwise to
a maximum torque of 20 in-lb, using a flat blade torque
screwdriver.

6. Install a filler panel over any empty slot, and secure in place at the

top and bottom by turning the captive thumbscrews clockwise to a
maximum torque of 20 in-lb, using a flat blade torque screwdriver.