31 lpal resistance calibration part v of v, Lpal – AMETEK ReFlex User Manual
Page 355
Active Loads
ReFlex Power™
7-80
M380056-01 Rev J
7.6.31
LPAL
R
ESISTANCE
C
ALIBRATION
P
ART
V
OF
V
This procedure calibrates the resistance for voltage gain 1 and current
range low. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Connect the
sense positive-to-positive lead and sense negative-to-negative lead. Also
connect the shunt and the DC current meter between the active load and
DC power supply
This procedure requires maximum voltage 500.0 V and maximum current
0.5 Amps.
1. Set the DC input voltage to 500 Volts.
2. Turn the active load on:
INP
3. Set the conductance count to 0x246:
CAL
4. After DC voltage stabilizes, enter the voltage:
CAL
5. After DC current stabilizes, enter the current:
CAL
6. Set the conductance count to 0xF9:
CAL