beautypg.com

31 lpal resistance calibration part v of v, Lpal – AMETEK ReFlex User Manual

Page 355

background image

Active Loads

ReFlex Power™

7-80

M380056-01 Rev J

7.6.31

LPAL

R

ESISTANCE

C

ALIBRATION

P

ART

V

OF

V

This procedure calibrates the resistance for voltage gain 1 and current
range low. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Connect the
sense positive-to-positive lead and sense negative-to-negative lead. Also
connect the shunt and the DC current meter between the active load and
DC power supply

This procedure requires maximum voltage 500.0 V and maximum current
0.5 Amps.

1. Set the DC input voltage to 500 Volts.

2. Turn the active load on:

INP:STATE 1

3. Set the conductance count to 0x246:

CAL:INP:COND:COUNTS 0x246

4. After DC voltage stabilizes, enter the voltage:

CAL:INP:VOLT 1

5. After DC current stabilizes, enter the current:

CAL:INP:CURR 1

6. Set the conductance count to 0xF9:

CAL:INP:COND:COUNTS 0xF9

This manual is related to the following products: