5 output current parameter calibration, Output, Current – AMETEK ReFlex User Manual

Page 272: Parameter, Calibration

background image

ReFlex Power™

DC Power Supplies

M380056-01 Rev K

6-69

*RST[n]

6.6.5

OUTPUT

CURRENT

PARAMETER

CALIBRATION

The following procedure calibrates parameters associated with the output current:
programming current, readback current, overcurrent protection (OCP), and readback of
local current. Output current calibration is performed with the unit operating in current-
mode and with the output short-circuited through a current shunt. The current is
determined by measuring the burden voltage and calculating the corresponding current.

1. Connect a current shunt across the output terminals, and connect the external

voltmeter to measure the burden voltage of the shunt.

2. Issue clear and reset commands:

*CLS[n]

*RST[n]

3. Set the OVP supervisory to non-tracking and at maximum range:


CAL[n]:OUTP:VOLT:PROT:COUNTS 4095

4. Set the OCP supervisory to non-tracking and at maximum range:


CAL[n]:OUTP:CURR:PROT:COUNTS 4095

5. Turn on the output:

OUTP[n]:ISOL on

OUTP[n]:SENS on

OUTP[n]:STAT on

6. Set the output voltage to full-scale:


CAL[n]:OUTP:VOLT:COUNTS 49152

7. Program the output current to the first value near 5% of full-scale,

(count = 32768*0.04611 =1511):

CAL[n]:OUTP:CURR:COUNTS 1511

8. Measure the output current with the external shunt. Enter this as

value-1 with the following command:

CAL[n]:OUTP:CURR:FIVEPOINT 1

9. Program the output current to the second value near 27.5% of full-scale,