AMETEK ReFlex User Manual
Page 139
AC Power Supplies
ReFlex Power™
4-54
M380056-01 Rev K
4.6.8
V
OLTAGE
ADC
O
FFSET
C
ALIBRATION FOR
V
OLTAGE
R
ANGE
L
OW
The following procedure calibrates offset of the voltage ADC for the voltage
range low.
1. Ensure that no load is connected to the output.
2. Issue clear and reset commands:
*CLS
*RST
3. Perform the voltage ADC offset calibration
CAL
4. Send the system error query command to make sure calibration was successful.
SYST:ERR?
5. If there is no error returned at the system error query command then save the
calibration result with the following commands
CAL
CAL
CAL
6. Optional Commands:
To query the voltage ADC offset:
CAL
7. Reset the unit
*RST