background image

ReFlex Power™

AC Power Supplies

M380056-01 Rev K

4-59

4.6.13

H

IGH

S

PEED

S

UPERVISOR

C

ALIBRATION FOR

V

OLTAGE

R

ANGE

H

IGH

The following procedure calibrates offset for the V-Local, I-Local and V-Remote

1. Ensure that no load is connected to the output.

2. Issue clear and reset commands:

*CLS
*RST

3. Set the voltage range of the power supply to high

SOUR:VOLT:RANGE 1

4. Perform the voltage ADC offset calibration

CAL:MEAS:OFFSET:CALC

5. Send the system error query command to make sure calibration was successful.

SYST:ERR?

6. If there is no error returned at the system error query command then save the

calibration result with the following commands

CAL:UNLOCK “6867”
CAL:STORE
CAL:LOCK

7. Optional Commands:

To query the offset of signal V-Local:

CAL:MEAS:VLOCAL:OFFSET ?

T

o query the offset of signal I-Local:

CAL:MEAS:ILOCAL:OFFSET ?

To query the offset of signal V-Remote:

CAL:MEAS:VREMOTE:OFFSET ?

8. Reset the unit

*RST