ReFlex Power™

Active Loads

M380056-01 Rev J

7-47

7.6.15

HPAL

R

ESISTANCE

C

ALIBRATION

P

ART

III

OF

V

The following procedure calibrates the resistance for voltage gain 4 and
current range high. Make sure that the shunt and the DC voltmeter are
connected between the positive and negative leads of the active load.
Connect the sense positive-to-positive lead and sense negative-to-negative
lead. Also connect the shunt and the DC current meter between the active
load and DC power supply.

This procedure requires maximum voltage 50.0 V and maximum current 2.0
Amps.

1. Set the DC input voltage to 50 Volts

2. Turn the active load on:

INP:STATE 1

3. Set the conductance count to 0x7FFF:

CAL:INP:COND:COUNTS 0x7FFF

4. After DC voltage stabilizes, enter the voltage:

CAL:INP:VOLT 1

5. After DC current stabilizes, enter the current:

CAL:INP:CURR 1

6. Set the conductance count to 0x3C3C:

CAL:INP:COND:COUNTS 0x3C3C

7. After DC voltage stabilizes, enter the voltage: