DC Power Supplies

ReFlex Power™

6-68

M380056-01 Rev K

CAL[n]:OUTP:VOLT:COUNTS 14202

9. Measure the output voltage with the external voltmeter. Enter this as value-2 with

the following command:

CAL[n]:OUTP:VOLT:FIVEPOINT 2 2

10. Program the output voltage to the third value near 50% of full-scale,

count = 48852*0.50+768 = 25194):

CAL[n]:OUTP:VOLT:COUNTS 25194

11. Measure the output voltage with the external voltmeter. Enter this as value-3 with

the following command:

CAL[n]:OUTP:VOLT:FIVEPOINT 3

12. Program the output voltage to the fourth value near 72.5% of full-scale,

(count = 48852*0.725+768 = 36185):

CAL[n]:OUTP:VOLT:COUNTS 36185

13. Measure the output voltage with the external voltmeter. Enter this as value-4 with

the following command:

CAL[n]:OUTP:VOLT:FIVEPOINT 4

14. Program the output voltage to the fifth value near 95% of full-scale,

(count = 48852*0.95+768 = 47177):

CAL[n]:OUTP:VOLT:COUNTS 47177

15. Measure the output voltage with the external voltmeter. Enter this as value-5 with

the following command:

CAL[n]:OUTP:VOLT:FIVEPOINT 5

16. Save the calibration results with the following commands:

CAL[n]:UNLOCK “6867”
CAL[n]:STORE
CAL[n]:LOCK


(The entered voltage data could be verified with the following query:

CAL[n]:OUTP:VOLT:FIVEPOINT?)

17. Turn off the output: