beautypg.com

7 environmental characteristics, Operating temperature, Operating altitude – AMETEK ReFlex User Manual

Page 287: Operating humidity, Operating shock, Operating vibration, Operation without fully populated mainframe, Non-operating temperature, Non-operating altitude, Non-operating shock

background image

Active Loads

ReFlex Power™

7-12

M380056-01 Rev J

7.3.7

E

NVIRONMENTAL

C

HARACTERISTICS

O

PERATING

T

EMPERATURE

Standard: 0 to 50 °C

Extended Range Option: -10 to 55 °C

O

PERATING

A

LTITUDE

Standard: 0 to 6,500 ft

Extended Range Option:

Low-Power Load assets: 0 to 15,000 feet

High-Power Load asset: 0 to 15,000 feet, with linear temperature
derating from 55

°C at 6,500 feet to 40°C at 15,000 feet, or with linear

power derating from 750W at 6,500 feet to 600W at 15,000 feet

O

PERATING

H

UMIDITY

95% non-condensing

O

PERATING

S

HOCK

Class 3 MIL-PRF-28800F

O

PERATING

V

IBRATION

Class 3 MIL-PRF-28800F

O

PERATION

W

ITHOUT

F

ULLY

P

OPULATED

M

AINFRAME

If Mainframe is not fully populated with modules, filler panels must be
installed in empty slots to maintain proper airflow and structural
integrity.

N

ON

-O

PERATING

T

EMPERATURE

-40 to 71 °C

N

ON

-O

PERATING

A

LTITUDE

0 to 50,000 ft

N

ON

-O

PERATING

S

HOCK

Class 3 MIL-PRF-28800F

N

ON

-O

PERATING

V

IBRATION

Class 3 MIL-PRF-28800F

This manual is related to the following products: