7 environmental characteristics, Operating temperature, Operating altitude – AMETEK ReFlex User Manual
Page 287: Operating humidity, Operating shock, Operating vibration, Operation without fully populated mainframe, Non-operating temperature, Non-operating altitude, Non-operating shock
Active Loads
ReFlex Power™
7-12
M380056-01 Rev J
7.3.7
E
NVIRONMENTAL
C
HARACTERISTICS
O
PERATING
T
EMPERATURE
Standard: 0 to 50 °C
Extended Range Option: -10 to 55 °C
O
PERATING
A
LTITUDE
Standard: 0 to 6,500 ft
Extended Range Option:
Low-Power Load assets: 0 to 15,000 feet
High-Power Load asset: 0 to 15,000 feet, with linear temperature
derating from 55
°C at 6,500 feet to 40°C at 15,000 feet, or with linear
power derating from 750W at 6,500 feet to 600W at 15,000 feet
O
PERATING
H
UMIDITY
95% non-condensing
O
PERATING
S
HOCK
Class 3 MIL-PRF-28800F
O
PERATING
V
IBRATION
Class 3 MIL-PRF-28800F
O
PERATION
W
ITHOUT
F
ULLY
P
OPULATED
M
AINFRAME
If Mainframe is not fully populated with modules, filler panels must be
installed in empty slots to maintain proper airflow and structural
integrity.
N
ON
-O
PERATING
T
EMPERATURE
-40 to 71 °C
N
ON
-O
PERATING
A
LTITUDE
0 to 50,000 ft
N
ON
-O
PERATING
S
HOCK
Class 3 MIL-PRF-28800F
N
ON
-O
PERATING
V
IBRATION
Class 3 MIL-PRF-28800F