21 lpal low range current calibration, Lpal – AMETEK ReFlex User Manual

Page 335

background image

Active Loads

ReFlex Power™

7-60

M380056-01 Rev J

7.6.21

LPAL

L

OW

R

ANGE

C

URRENT

C

ALIBRATION

This procedure calibrates the current for the low range. Make sure that the
shunt and the DC current meter are connected between the active load and
DC power supply.

This procedure requires maximum voltage 5.0 V and maximum current 0.5
Amps.

1. Set the DC input voltage to 5 Volts

2. Turn on the active load

INP:STATE 1

3. Set the DAC count to 20% of its full range

CAL:INP:CURR:COUNTS 16000

4. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL:INP:CURR1

5. Set the DAC count to 80% of its full range

CAL:INP:CURR:COUNTS 53000

6. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL:INP:CURR2