Output voltage/current regulation – AMETEK ReFlex User Manual

Page 116

background image

ReFlex Power™

AC Power Supplies

M380056-01 Rev K

4-31

through software. In a multi-phase group the load could be connected in wye or
delta configuration, but the module outputs must be wye connected.

CAUTION

Do not connect the AC modules in a delta configuration.

Modules must be set to the same output range when multi-phase connected.

O

UTPUT

V

OLTAGE

/

CURRENT

R

EGULATION

The AC power supply module is capable of operating while regulating either the
output voltage, constant–voltage mode (CV), or the output current, constant–
current mode (CC). Which parameter would be regulated is dependent on the
programmed settings of the output voltage and current in relation to the
resistance of the load. If the load resistance is greater than the voltage setting
divided by the current setting, the output voltage would be regulated in voltage-
mode. If the load resistance is less than the voltage setting divided by the
current setting, the output current would be regulated in current-mode. The
crossover point is when the load resistance equals the voltage setting divided
by the current setting. The AC power supply can automatically crossover
between the two modes of operation in response to load demands. Refer to
Figure 4-9 for a diagram showing the operating regions for constant-voltage
and constant-current.

V

Constant-Voltage
Regulation

Constant-Current
Regulation

Set

I

Set

Output

Current

Output

Voltage

R >

V

I

Set

Set

R <

V

I

Set

Set

R =

V

I

Set

Set

Where, R = Load Resistance

= Output Voltage Setting

= Output Current Setting

ISet

VSet

Figure 4-9. CV and CC Regulation