5 hpal high range current calibration, Hpal – AMETEK ReFlex User Manual
Page 309
Active Loads
ReFlex Power™
7-34
M380056-01 Rev J
7.6.5
HPAL
H
IGH
R
ANGE
C
URRENT
C
ALIBRATION
This procedure calibrates the input current for current range high. This is a 5-
point calibration. Make sure that shunt and the DC current meter are connected
between the active load and DC power supply.
This procedure requires maximum voltage 5.0V and maximum current 30Amps.
1. Set the DC input voltage to 5 Volts.
2. Turn on the active load:
INP
3. Set the DAC count to 5% of its full range:
CAL
4. Wait for the DC current to stabilize. After the current reading is stabilized,
enter the current using following command:
CAL
5. Set the DAC count to 27.5% of its full range:
CAL
6. Wait for the DC current to stabilize. After the current reading is stabilized,
enter the current using following command:
CAL
7. Set the DAC count to 50.0% of its full range:
CAL
8. Wait for the DC current to stabilize. After the current reading is stabilized,
enter the current using following command:
CAL
9. Set the DAC count to 72.5% of its full range:
CAL
10. Wait for the DC current to stabilize. After the current reading is stabilized,
enter the current using following command:
CAL
11. Set the DAC count to 95% of its full range
CAL