AMETEK ReFlex User Manual
Page 168

ReFlex Power™
AC Power Supplies
M380056-01 Rev K
4-83
14. Set the output frequency to 250Hz
SOUR
15. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current:
CAL
16. Set the output frequency to 440Hz
SOUR
17. Wait for AC current to stabilize. Measure the output current using the AC current
meter
CAL
18. Set the output frequency to 630Hz:
SOUR
19. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current:
CAL
20. Set the output frequency to 820Hz:
SOUR
21. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
22. Set the output frequency to 1010Hz
SOUR
23. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
24. Set the output frequency to 1200Hz:
SOUR
25. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current: