Active Loads

ReFlex Power™

7-50

M380056-01 Rev J

7. After DC voltage stabilizes, enter the voltage:

CAL:INP:VOLT 2

8. After DC current stabilizes, enter the current:

CAL:INP:CURR 2

9. Send the calibration command to calculate the gain and offset:

CAL:INP:COND:CALC

10. Send the system error query command to make sure calibration was

successful:

SYST:ERR?

11. If there is no error returned at the system error query command,

then save the calibration result as follows:

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

12. Reset the unit:

*RST