AMETEK ReFlex User Manual
Page 310

ReFlex Power™
Active Loads
M380056-01 Rev J
7-35
12. Wait for the DC current to stabilize. After the current reading is stabilized,
enter the current using following command:
CAL
13. Set DAC counts to zero:
CAL12:INP:CURR:COUNTS 0
14. Send calibration command to make sure that all the DAC counts and
current values are valid:
CAL
15. Send the system error query command to make sure calibration was
successful:
SYST:ERR?
16. If there is no error returned at the system error query command then save
the calibration result with the following commands:
CAL
CAL
CAL
17. Reset the unit:
*RST