7.6.19

LPAL

C

AL

P

OT

C

ALIBRATION

This procedure calibrates cal-pot. This calibration is done with the DC
current meter connected between the DC power supply and the Active
Load.

This procedure requires maximum voltage 4.5V and maximum current
15Amps.

1. Connect the shunt between the DC power supply and the active

load. Connect the DC current meter across the shunt.

2. Issue clear and reset commands:

*CLS

*RST

3. Set the DC power supply to 4.5 Volts

4. This begins the CalPots calibration sequence. The Load will be set

to STATE 1; relays closed and set to 0 current, the 0 current offset
value will be calculated.

CAL:CALPOTS:GAIN:START

5. Program to draw maximum current

CAL:INP:CURR:COUNTS 0xFFFF

6. Adjust the current limit on the DC Power supply to just under the

limit 13.7 Amps. Use the Current Shunt and Voltmeter to adjust and
measure the current. When the current reading is stable, enter the
current using following command

CAL:CALPOTS:GAIN:CURR