16 ldf calibration for voltage range low – AMETEK ReFlex User Manual

Page 150

background image

ReFlex Power™

AC Power Supplies

M380056-01 Rev K

4-65

4.6.16

LDF

C

ALIBRATION FOR

V

OLTAGE

R

ANGE

L

OW

The following procedure calibrates the Line drop fault supervisory.

1. Ensure that no load is connected to the output.

2. Issue clear and reset commands:

*CLS
*RST

3. Set the voltage range of the power supply to low

SOUR:VOLT:RANGE 0

4. Turn the unit on

OUTP:STATE 1

5. Program the output voltage to count 17000

CAL:OUTP:VOLT:COUNTS 17000

6. Send the calibration command to start the internal calibration

CAL:OUTP:LDF:CALC

7. Turn the unit off

OUTP:STATE 0

8. Send the system error query command to make sure calibration was successful.

SYST:ERR?

9. If there is no error returned at the system error query command then save the

calibration result with the following command

CAL:UNLOCK “6867”
CAL:STORE
CAL:LOCK

10. Optional Commands:

Query the voltage calibration points:

CAL:OUTP:LDF:GAIN ?

11. Reset the unit

*RST