Standalone configuration operation check – AMETEK ReFlex User Manual
Page 126
ReFlex Power™
AC Power Supplies
M380056-01 Rev K
4-41
the slot of the ReFlex Power™ Mainframe where the module is located,
counting left to right, 1 through 12.
The ReFlex Power™ Controller module should respond with its identification
string.
Verify communications with the ReFlex Power™ AC power supply module by
issuing an SCPI *IDN[n]?, where “n” is a channel number that
corresponds to the slot of the ReFlex Power™ Mainframe where the
module is located, counting left to right, 1 through 12.
The AC power supply module should respond with its identification string.
The default settings of the module will not be affected.
S
TANDALONE
C
ONFIGURATION
O
PERATION
C
HECK
Connect the output for standalone operation per the section, Standalone
Configuration.
Connect appropriate instrumentation for measuring the output voltage and
current, as well as a load.
Perform the power-on verification test per the section, Power-On Check.
Turn on the isolation and remote sense relays, turn on the output converter,
and program full scale voltage and current with the following commands:
*RST[n]
OUTP[n]:STATE 1
SOUR[n]:VOLT
SOUR[n]:CURR
Measure the output voltage and current with the external instruments.
Ensure that the actual values of the output parameters are within
specifications of the programmed values.
Query the values of the output parameters with the following commands:
MEAS[n]:VOLT?
MEAS[n]:CURR?
Ensure that the readback values of the output parameters are within
specifications of the actual values.
Set the frequency to 400 Hz: SOUR[n]:FREQ 400
Close the auxiliary relay: OUTP[n]:AUXR 1
Verify the proper output voltage at the AUX output (should equal 22% of
VOLT fval as entered in Step 4 above).