10 hpal voltage calibration – part iii of v, Hpal – AMETEK ReFlex User Manual

Page 315

background image

Active Loads

ReFlex Power™

7-40

M380056-01 Rev J

7.6.10

HPAL

V

OLTAGE

C

ALIBRATION

P

ART

III

OF

V

The following procedure calibrates the voltage for voltage gain 4 and current
range high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Also connect the
sense positive-to-positive lead and sense negative-to-negative lead.

This procedure requires maximum voltage 218V and will draw maximum current
3.0Amps.

1. Turn the active load on

INP:STATE 1

2. Set the conductance count to 0x3333

CAL:INP:COND:COUNTS 0x3333

3. Set the input DC voltage to 54 Volts. After the input DC voltage stabilizes

enter the first exact voltage using following command

CAL:INP:VOLT 1

4. Set the input DC voltage to 218 Volts. After the input DC voltage

stabilizes enter the second exact voltage using following command

CAL:INP:VOLT 2

5. Send the calibration command to calculate the gain and offset:

CAL:MEAS:OFFSET:CALC

6. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

7. If there is no error returned at the system error query command then save

the calibration result as follows:

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

8. Reset the unit:

*RST