13. Wait for AC voltage to stabilize. Measure the output voltage using the AC

voltmeter. Enter the measured voltage using following command

CAL:OUTP:VOLT:FIVEPOINT2

14. Program the output voltage to 50% of the full scale voltage

CAL:OUTP:VOLT:COUNTS 9649

15. Wait for AC voltage to stabilize. Measure the output voltage using the AC

voltmeter. Enter the measured voltage using following command

CAL:OUTP:VOLT:FIVEPOINT3

16. Program the output voltage to 72.5% of the full scale voltage

CAL:OUTP:VOLT:COUNTS 13990

17. Wait for AC voltage to stabilize. Measure the output voltage using the AC

voltmeter. Enter the measured voltage using following command

CAL:OUTP:VOLT:FIVEPOINT4

18. Program the output voltage to 95% of the full scale voltage

CAL:OUTP:VOLT:COUNTS 18332

19. Wait for AC voltage to stabilize. Measure the output voltage using the AC

voltmeter. Enter the measured voltage using following command

CAL:OUTP:VOLT:FIVEPOINT5

20. Perform the voltage calibration using following command

CAL:OUTP:VOLT:CALC

21. Send the system error query command to make sure calibration was successful.

SYST:ERR?

22. If there is no error returned at the system error query command then save the

calibration result with the following command

CAL:UNLOCK “6867”
CAL:STORE
CAL:LOCK

23. Optional Commands:

Query the voltage calibration points:

CAL:OUTP:VOLT:FIVEPOINT ?

24. Reset the unit