AMETEK ReFlex User Manual
Page 146
ReFlex Power™
AC Power Supplies
M380056-01 Rev K
4-61
13. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL
14. Program the output voltage to 50% of the full scale voltage
CAL
15. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL
16. Program the output voltage to 72.5% of the full scale voltage
CAL
17. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL
18. Program the output voltage to 95% of the full scale voltage
CAL
19. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL
20. Perform the voltage calibration using following command
CAL
21. Send the system error query command to make sure calibration was successful.
SYST:ERR?
22. If there is no error returned at the system error query command then save the
calibration result with the following command
CAL
CAL
CAL
23. Optional Commands:
Query the voltage calibration points:
CAL
24. Reset the unit