AMETEK ReFlex User Manual
Page 331

Active Loads
ReFlex Power™
7-56
M380056-01 Rev J
7. This will use the current value entered to adjust the Gain Pot. The
adjustment will be calculated, ending the procedure and the setting
the load to STATE 0
CAL
8. Send the system error query command to make sure calibration was
successful.
SYST:ERR?
9. If there is no error returned at the system error query command then
save the calibration result with the following command
CAL
CAL
CAL
10. Optional Commands:
To query the offset pot set value: CAL
To query the gain pot set value: CAL
11. Reset the unit
*RST