Active Loads

ReFlex Power™

7-56

M380056-01 Rev J

7. This will use the current value entered to adjust the Gain Pot. The

adjustment will be calculated, ending the procedure and the setting
the load to STATE 0

CAL:CALPOTS:GAIN:CALC

8. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

9. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

10. Optional Commands:

To query the offset pot set value: CAL:CALPOTS:OFFSET?

To query the gain pot set value: CAL:CALPOTS:GAIN?

11. Reset the unit

*RST